General
The SamplingSystem is designed for filtration and conveyance for analytic measures in hot conditions with small volumes. All exhaust gas exposed parts of the system are made of stainless steel, nickelplated aluminium or Teflon and are incorporated in a heated oven.
Using an external temperature controller, the oven is controlled to a maximum of 200°C. The temperature sensor is Pt100 and the heater is secured by a 225°C temperature safety switch.
At the front you can see filter, gas access, plug for heated incoming lead, main switch and operating terminal, at rear the gas outgoing, mains plug, plug for heated outgoing lead and computer infaces.
The SamplingSystem is designed and optimized to work with our IMR mass spectrometers.

Functional Diagram

Typical Applications
- Development of combustion engines (diesel and gasoline)
- Catalysts development
- Fast measurement of non regulated hydrocarbons
- Ammonia measurements on DeNOx catalysts
- NOx measurements on diesel engines
Spezifications
Technical Data | Value | Technical Data | Value |
Sample tray positions | 70 | Power | 230 V/50 Hz or 110 V/60 Hz 750 W |
Heating oven positions | 12 | Dimension (HxWxD) | 56 x 46 x 64 cm |
Temperature | up to 230 °C | Weight | 47 kg |
Transfer line | deactivated |
Reference Customers (excerpt)